Products
 
   
   
   
   
   
   
   
   
   
   
   
   
   
   
   
 

Defibrillator

 
   
   
   
 
 
Oxy Pulse/Pulse Oximeter :
Versatile, Simple & Economical
Compact, Easy to use
Large LCD
Variable Pulse Strength Beeper
 10 Segment Pulse Strength Bar
• Patient Isolation as per IEC60601-1-1 standards
• Universal Patient Application
• Trend with Printer Interface
• Separate Audio for Alarm & Pulse Detection
• Extensive Self Diagnostic
  SPECIFICATIONS - DISPLAY
SCREEN
:
LCD Screen of 130mm X 38mm active area
NUMERIC Display
:
High brightness Seven Segment Display
PARAMETERS
:
Pleth Waveform, SpO2%, Pulse Rate (BPM), Bar graph display
SWEEP SPEED
:
12. 5, 25 and 5Omm/sec
BAND WIDTH
:
0.1 to 35Hz
CONTROLS
:


Freeze, Sweep Speed (12.5/25/50mm/sec), Alarm Settings, Trend Display Settings, LCD Brightness Control
 
   % SpO2
 
READ OUT
:
0 TO 100%
ACCURACY
:
:
±2% at 70 TO 100% SaO2
±3% at 50 TO 69% SaO2
PLESTHYMOGRAPH
:
Displayed in LCD and 15 Bit Bar Graph
WAVE GAIN
:
Auto Gained
TECHNOLOGY
:
Digital Auto Corr Technology for taking care of Motion Artifact
  PULSE
 
READ OUT
:
30 TO 250BPM
ACCURACY PULSE
:
±3%±1 BPM
AVERAGING TECHNIQUE
:
Moving Average Window indication in the LCD Panel
QRS INDICATION
:
Indication in the LCD Panel
 
  INDICATIONS
 
READ OUT
:
Alarm, Sensor Off, Finger Off, Pulse Strength,
Real Time Clock
:
ALARMS
:
ALARM LIMIT SETTINGS
:
Independent of all Parameters Last Set values are stored in Non-Volatile Memory
ALARM ACCURACY
:
+/-1
AUDIO/VISUAL
:
Low & High Alarm Violation, Continuous Beep for Alarm Violation
   
  TREND
 
READ OUT
:
Both Numeric and Waveform Type for both parameters Independent of Current Information
DURATION
:
24 hours selectable from 30 minutes
  OPTIONS
HARDWARE
:
RS 232 Port for hookup with CMS, Printer Port for Printing Trend Data, Analog Data Out
SOFTWARE
:
0 to 30 Mins Period can be selected for Manual printing.
SENSORS
:
All Types of Universal, Adult, Neonatal both Reusable and Disposable Sensor - Range of Probes available
  BATTERY
 
TYPE
:
Lead Acid, 1 2V, 2.0aH
CHARGE TIME
:
16 Hour
OPERATION TIME
:
3 Hrs.
  GENERAL
DIMENSION
:
100mm(H) x 300mm(L) x 240mm(W)
WEIGHT
:
6.9 Kgs.
POWER INPUT
:
230V ± 30% AC, 50Hz,
CONSUMPTION
:
1OOVA
EARTH LEAKAGE CURRENT
:
<I00 micro A
ELECTRICAL SAFETY
:
Meets IEC 60601-1 standards
PROTECTION
:
Against Cautery and Defibrillation
OPERATING TEMPERATURE
:
+ 5 deg C to +40 deg C at 95% RH-Non condensing
 
     